ILT-710-V.5
Exhibitor
ESA Elektroschaltanlagen Grimma GmbH
The ILT-710-V.5 is a multifunctional monitoring device for medical IT systems, specifically designed for use in accordance with the IEC 60364-7-710 and DIN VDE 0100-710 standards. As an integral component of the HospEC® system, it combines several essential monitoring functions in one compact module.
Main Monitoring Functions
Insulation monitoring: The device continuously monitors the insulation resistance of single-phase medical IT systems (AC 110–230 V).
Load current monitoring: It measures the load current of single-phase IT system transformers using a special external current transformer (type ILT-W).
Temperature monitoring: The transformer's temperature is reliably monitored using PTC thermistors or break contacts.
Insulation Fault Detection (IFS)
The ILT-710-V.5 features an integrated test signal generator.
In combination with at least one insulation fault detection device (e.g., the IFS-710-W6), it can be expanded into a complete insulation fault detection system in accordance with IEC 61557-9.
If a fault occurs, a test signal is fed in, which allows the defective outgoing circuit to be exactly localized and its name (e.g., fuse designation) to be shown on the display.
Operation and Display
The device is equipped with a full-graphics LCD display (128x64 pixels with white backlighting), which clearly shows all operating and fault messages in plain text.
A multi-coloured "Status" LED indicates the current operating status at a glance (e.g., green for normal operation, flashing red for device errors, yellow when response values are exceeded).
User-friendly parameterization is carried out via four control keys.
A non-volatile message memory logs up to 200 events with a date and time stamp.
Communication and Interfaces
All messages are transmitted to peripheral operating and annunciator terminals via the integrated CAN bus (CAN 2.0, Protocol: ESA-CAN).
The device supports the Modbus RTU protocol (as a slave) via an RS485 interface.
Two configurable connections for fault signalling relays enable easy integration into the building control technology (BSCS).
Safety and Standard Compliance
The ILT-710-V.5 performs continuous automatic self-monitoring, including a PE and KE connection test every 30 minutes, as well as checking the transformer connections for interruptions and short circuits.
An integrated test alarm function monitors the operator's obligation to carry out periodic inspections (e.g., according to DIN VDE 0100-710).
The device is CB-certified by VDE and meets international standards such as IEC 61557-1, IEC 61557-8, IEC 61557-9, and IEC 61010-1.